Question:What does it mean when F34 displays on the Leland Legacy Sample Pump?
Answer:Normally, F34 displays on the Leland Legacy pump when the temperature sensor on the control board is broken. Contact SKC for repair service.

SKC Catalog Number:100-3002, 100-3, 761-200, 761-203B, 761-20
Category:Particulate Sampling
Sub-category:Particle Size-selective Sampling
Question:Can the Leland Legacy Sample Pump be used with the PEM for 24-hour sampling?
Answer:Yes. The Leland Legacy Sample Pump may be used with the PEM at 10 L/min for 24-hour sampling when LOW back pressure filters, such as PTFE filter 225-1709, are used. The overall back pressure across the PEM/filter should not exceed 15 inches of water.

Note: Back pressure on PTFE filters can vary within the same lot.

Calibration
The PEM high flow sampling train creates a greater load for the constant flow Leland Legacy Sample Pump. To ensure successful calibration, use a recommended low back pressure filter, the calibration cap designed for the PEM (not a calibration jar), and consider that calibration devices such as calibrators and rotameters introduce additional back pressure to the total back pressure of the train.

SKC Catalog Number:100-3002, 210-5001, 210-5002
Category:Sample Pumps
Sub-category:Sample Pumps
Question:How long can a Leland Legacy or AirChek XR5000 battery pack be kept on the 223-241 charger?
Answer:The battery pack may be kept on the 223-241 charger indefinitely.

 