Question:224-52 samplers: How are they used in low flow mode?
Answer:Low flow applications on the 224-52 require a constant pressure controller (224-26-CPC) and an adjustable low flow holder (224-26 Series):
  1. Turn the pump on and adjust the flow to approximately 1.5 L/min.
  2. Connect the outlet side of the constant pressure controller (CPC) to the pump intake using a short length of tubing.
  3. Connect the inlet side of the CPC to the adjustable low flow holder
  4. Insert an opened tube in the adjustable low flow holder's rubber sleeve and connect a flowmeter to the exposed end of the tube.
  5. Turn the pump on and turn the adjustment screw ON THE ADJUSTABLE FLOW HOLDER until the flowmeter displays the desired flow rate.