  Home » Catalog » Pumps/Calibrators/Accessories » Calibrators and Accessories » Calibrators & Rotameters » chek-mate Calibrator, 0.75 to 5 L/min

chek-mate Calibrator, 0.75 to 5 L/min

Orderable Items The Proof is in the Performance! With a volumetric accuracy of 1%, and a price $500 less than a Defender, the chek-mate is the smart choice for air sample pump calibration!

500 dollar savings burst
    • 0.75 to 5 L/min flow measurement range

    • Volumetric accuracy 1% of reading
    • Displays continuous real-time volumetric flow rate readings in L/min
      • Moving average of the measured flow rate

    • <10 minutes stabilization to ambient temperature

    • Calibration laboratory accuracy certified by NIST and backed by ISO 17025 accreditation
      • NIST traceable calibration certificate supplied

    • Sleek, portable design
      • Only 8.5 ounces (242 grams)

    • No moving parts, use in any orientation

    • One-button operation

    • 9-volt alkaline battery operation
      • Approximately 30 hours battery life
      • Auto shut-off feature conserves battery

    • Calibration and Certification Services








Performance Profile

Flow Measurement Range 0.75 to 5 L/min (volumetric)
Display Resolution 0.001 L/min
Accuracy ± 1% of volumetric reading

9-V alkaline primary (disposable) PP3 / 6LR61 / 1604A or equivalent, or
8.4V NiMH secondary (rechargeable) PP3 / 6HR61 / 8.4H5 or equivalent

Automatic power off timer – 15 minutes

Operating Temperature Range 32 to 104° F (0 to 40° C)

Operating Barometric Pressure

700 mbar to 1090 mbar (20.7 inch Hg to 32.2 inch Hg)

Operating Altitude

Sea level to approximately 10,000 ft (3050m) above sea level

IP Rating (enclosure)


Overall Dimensions
(including hose tail)

7 x 3.3 x 1.3 in (17.8 x 8.4 x 3.3 cm)


8.5 ounces (242 gm)


CE Marked


Chek Mate Sample Train

How chek-mate Works

The SKC chek-mate calibrator is a highly reliable orifice-type airflow meter that incorporates sensors to measure pressure drop across a fixed orifice. Sensors automatically correct for changes in air temperature and atmospheric pressure to provide accurate volumetric flow rate readings. Volumetric accuracy is 1% of reading and highly repeatable.