Film Flowmeters
Features
- Primary standards
- Simple to use
- Sample contacts only glass and stainless steel
- Calibration lines encircle the tube for easy reading
Principle of Operation
The operating principle of the SKC film flowmeter is simple: the sampling pump
to be calibrated is attached to the top of a volumetric glass tube containing
a small amount of liquid soap. As the pump draws air through the tube, a flat
soap bubble is interposed in the flow path. While the air flow causes the soap
film to move from one volume mark to another, the travel time is measured with
a stopwatch. The flow rate can then be directly calculated using the travel time
and the known tube volume, or it can be read from the flow chart. The test is
quick and easy to perform. With a film flowmeter, flow rates are based upon the
inherent, unchanging properties of the measuring apparatus.
Model 303 Portable Flowmeter Kit
Range: 5 to 500 ml/min
The SKC 303 Portable Flowmeter Kit contains all of the equipment, except
a stopwatch, required for calibrating low flow personal sample pumps and
detector tube pumps. The complete kit weighs only 3 pounds (1.4 kg) and includes
a 100
ml buret flowmeter, film solution, operating instructions, flow charts,
and a removable carry case. Calibration lines for 90, 95, 105, and 110 ml are
also on the buret allowing for the calibration of detector tube pumps
 |
|
 |
303 Flowmeter |
|
Pump Calibration Train with 303 Flowmeter |
Model 303 Operating Instructions
(PDF)
Instrucciones
de Operación
Catálogo
No. 303 (PDF)
Model 311 Laboratory Film Flowmeter Kit
Range: 300 to 3000 ml/min
 |
|
|
| 311 Flowmeter |
|
Closeup of 311 Flowmeter side arm for pressure sources |
The SKC 311 Laboratory Film Flowmeter Kit consists of a glass tube with
red calibration lines at specified volumes, a squeeze bulb to contain sufficient
soap film
solution
for flow measurement, an adapter to accommodate either vacuum or pressure sources,
a stopper and connecting tubing, a tripod stand, film solution, and operating
instructions. The SKC 311 Laboratory Film Flowmeter Kit provides quick and
accurate flow determinations at specified flow rates.
Model 311
Operating
Instructions
(PDF)
Instrucciones de Operación
Catálogo
No. 311 Series (PDF)
About Built-in Flowmeters and Rotameters
The ball-type flowmeters and rotameters built into most air sampling pumps are
primarily intended to serve as flow indicators and are, therefore, of low accuracy.
These flowmeters as well as rotameters must be calibrated against a primary standard
such as a film flowmeter or electronic calibrator (
Defender)
to verify accuracy.
Precision Digital Stopwatch
An accurate timer is required with a film calibrator. The SKC stopwatch is a
rugged, hand-held digital stopwatch with high-precision quartz crystal movement
and solid-state circuitry powered by three AA alkaline batteries. The digital
readout is to 1/100 second with a total elapsed time of 60 minutes.
Ordering Information Online
Ordering
| Calibrator Size/Description |
Cat. No. |
| Portable Flowmeter Kit, 5 to 500 ml/min flowrange includes
precision glass buret, film solution, operating instructions, and carry
case |
 |
303 |
| Laboratory Film Flowmeter Kit, 300 to 3000 ml/min flowrange includes
precision glass buret, film solution, operating instructions, adapter
for vacuum or pressure,
and
aluminum tripod stand |
 |
311-1000 |
| Calibrating Accessories |
Cat. No. |
| Film Solution,1 pint (473 ml) |
302-4011 |
Precision Digital Stopwatch
Precision quartz crystal movement, digital readout to 1/100 second, powered
by 3 AA batteries |
303-01-1 |
Tubing Page
Soap Film
Flowmeters Spec. Sheet (PDF)
Model 303 Operating Instructions
(PDF)
Instrucciones de Operación
Catálogo
No. 303 (PDF)
Model 311
Operating Instructions
(PDF)
Instrucciones de Operación
Catálogo
No. 311 Series (PDF)
This page was last modified April 20, 2009