QuickTake 30 Sample Pump

Take High Flow on the Go!

QuickTake 30 Air Sampling Pump QuickTake 30 Air Sampling Pump
QuickTake 30 Air Sampling Pump QuickTake 30 Air Sampling Pump
  • 10 to 30 L/min constant flow sample pump
    • Maintains set flow throughout a sampling period for sample integrity
    • Ideal for use with BioStage Impactors, BioCassette, spore trap cassettes such as the VersaTrap, asbestos cassettes, wall cavity samplers, or other samplers that require flows to 30 L/min

  • Long run times on one Li-Ion battery charge
    • AC charger/adapter for extended sampling
    • Reliable, rechargeable lithium-ion battery pack

  • Easy-to-use programmable timer for unattended sampling
    • Select sample time presets or customize timer from 1 to 999 minutes
    • Continuous run with manual shut-off
    • Intermittent sampling
    • Large, easy-to-use keypad

  • Compact and portable
    • 9.3 x 8.4 x 3.5 in (23.6 x 21.3 x 8.9 cm)
    • Weighs only 4.8 lbs (2.2 kg)
    • Handle for easy portability

  • Versatile sampling accessories
    • Wand for sampling in ductwork and other hard-to-reach areas
    • Mounting bracket for BioStage allows impactor to be mounted directly to pump

  • Easy-to-use
    • Simple no-tool calibration using any reliable calibrator
    • End-of-cycle visual and audible alarms
    • Handy front-mounted intake for fast and simple media setup

  • High-contrast digital display


QuickTake® 30 Operating Instructions (PDF)

QuickTake 30 Spec. Sheet (PDF)

Technical Note: SKC Sample Pump Battery Pack Characteristics and Maintenance (PDF)


Performance Profile

Flow range
10 to 30 L/min
Flow compensation accuracy
± 5% of set flow
Compensating flow back pressure range

10 L/min at 90 inches water back pressure
20 L/min at 50 inches water back pressure
30 L/min at 15 inches water back pressure
Typical Back Pressure of Sampling Media (inches water)
Flow Rate
10.0
12.0
15.0
20.0
Filter/Pore Size (µm)
 
25-mm MCE/0.8
65
80
103
148
25-mm MCE/0.45
138
172
> 200
37-mm MCE/0.8
22
28
36
51
37-mm PVC/5.0
9
11
15
21
Compare the information in this table to compensation range to determine appropriate applications.
Power
Rechargeable 7.2-volt lithium-ion (Li-Ion) battery or battery with AC charger/adapter (100-240 V)
Battery recharge time
Approximately 5 hrs
Typical cumulative run time

Spore Trap (VersaTrap): 5 hrs at 15 L/min (battery only)
BioStage viable cascade impactor: 4 hrs at 28.3 L/min (battery only)
25-mm viable cascade impactor: 0.8-µm MCE filter: 1.5 hrs at 10 L/min (battery only)
25-mm, 0.8-µm MCE filter: 2.5 hrs at 10 L/min (fully charged battery plus AC charger/adapter)
25-mm, 1.2-µm MCE filter: For maximum air volumes during asbestos clearance sampling, use QuickTake 30 with a 25-mm, 1.2-µm MCE filter at 10 L/min with AC power to achieve 9+ hrs of run time for a total air volume of 5400 L.
37-mm, 0.8-µm MCE filter: 14 hrs at 10 L/min (battery only)

Note: The AC charger/adapter can be used to extend battery run time, but it is not a battery eliminator. Therefore, it will not provide indefinite run times.

Temperature
Operating: 32 to 113 F (0 to 45 C)
Storage: -4 to 113 F (-20 to 45 C)
Charging:
32 to 113 F (0 to 45 C)
Run time features
User-selectable features, user-adjustable presets. See Operating Instructions.
Preset timed runs: 1, 2, 5, 10, or 15 min
Manually set continuous run with manual shut-off: 1 to 999 min (repeats 1 to 999-min runs indefinitely until user stops pump or power supply is depleted)
Intermittent sampling:
User-set sample periods and sample delays. See Operating Instructions.
Flow fault features
If the pump is unable to compensate due to excessive back pressure, the pump will go into flow fault mode. Flow fault features are enabled as the default setting. Flow fault may be disabled at the user’s option. See Operating Instructions.
Flow fault: Flow fault displays immediately. Pump shuts off after 5 seconds
Fault restart: Pump attempts restart every 10 seconds up to five times
Media compatibility
BioStage viable cascade impactors, spore trap cassettes, wall cavity samplers, and other samplers that require flows to 30 L/min
Housing
ABS plastic
Dimensions
9.3 x 8.4 x 3.5 in (23.6 x 21.3 x 8.9 cm)
Weight
4.8 lbs (2.2 kg)


Use in non-hazardous locations only. Not UL Listed for intrinsic safety.
Protect sample pump from weather when in use outdoors.


Results obtained using a new pump and new fully charged battery. Pump performance may vary.

The SKC QuickTake 30 constant flow diaphragm pump maintains a set flow rate throughout the sampling period, compensating for the inherent back pressure created by sampling media and for any increase in back pressure as sample collects on the media. Unlike similar pumps, the QuickTake 30:

  • Maintains true constant flow

  • Can be calibrated with any primary standard calibrator for valid calibration

  • Does not require a special rotameter or hot-wire anemometer for calibration

  • Can be used with remote sampling wand and media that creates back pressure such as wall cavity samplers



Ideal for use with the BioStage      

BioStage Impactor attached to QuickTake 30 Sample Pump
QuickTake 30 with BioStage on mounting bracket

It's high flow sampling on the go! The portable SKC QuickTake 30 Sample Pump provides fast, easy sampling with the SKC BioStage Viable Cascade Impactor. Use the simple mounting bracket accessory to attach the BioStage directly to the QuickTake 30 pump. There is no need for bulky accessories or line voltage. Sampling with an impactor couldn't be easier!

QuickTake 30 228-9530 with VersaTrap 225-9820 cassette on sampling wand 228-9521
QuickTake 30 with VersaTrap on sampling wand

 

QuickTake 30 228-9530 with VersaTrap 225-9820 cassette

QuickTake 30 with VersaTrap


Ordering Information          

Description
Cat. No.

QuickTake 30 Sample Pump* includes battery pack and cassette and tubing adapters, requires 223-245 charger accessory below

Use in non-hazardous locations only. Not UL Listed for intrinsic safety.

QuickTake 30 Sample Pump and Charger* includes one pump with battery pack, AC charger/adapter (100-240 V), cassette and tubing adapters, and tubing

Use in non-hazardous locations only. Not UL Listed for intrinsic safety.

QuickTake 30 Sample Pump, Rotameter, and Charger* includes one pump with battery pack, AC charger/adapter (100-240 V), cassette and tubing adapters, rotameter, and tubing

Use in non-hazardous locations only. Not UL Listed for intrinsic safety.

QuickTake 30 BioStage Pump Kit* includes pump and charger as described above, Standard BioStage Impactor, mounting bracket with inlet adapter, calibration adapter, rotameter, and deluxe carry case

Use in non-hazardous locations only. Not UL Listed for intrinsic safety.

Certificate of Compliance, optional, must be ordered when ordering pump
Replacement Parts
Cat. No.
Replacement Inlet Filters, pk/50
Replacement Filter/O-ring set includes 5 filters and 1 0-ring
Stem Tubing Adapter, barbed tubing adapter for use with tubing and a standard filter cassette, pk/2
Cassette Adapter
Reducing Adapter for Tubing, 3/8 in to 1/4 in, pk/2
Lithium Ion Replacement Battery Pack* for QuickTake 30 Series pumps
Replacement Stack for QuickTake 30 Pump
Accessories
Cat. No.
Worldwide Charger/Adapter
Defender Primary Calibrator, 300 ml/min to 30 L/min, includes lead-acid battery, multi-plug charger (100-240 V), Optimizer 110 Software, and 1-meter serial cable
Sampling Wand, telescopes to 34 in (0.9 M) for ductwork and hard-to-reach locations, includes 6 ft (1.8 m) of tubing
Tripod Stand, telescopes to 5 ft (1.5 m) for mounting sample media or QuickTake 15 (QuickTake 30 pump will not mount on this stand)
Mounting Bracket for BioStage Impactor includes inlet adapter
Rotameter for QuickTake 15 and 30, to check approximate flow rate
Tygon Tubing, 3/8-in ID

10 ft
50 ft
Note:
The AC charger/adapter can be used to extend battery run time, but it is not a battery eliminator. Therefore, it will not provide indefinite run times.
*

Li-Ion Battery Shipment
Rechargeable, lithium-ion batteries for use with SKC sampling pumps have been tested in accordance with the UN Manual of Tests and Criteria and are designated as UN3091.

For air shipments:
Rechargeable lithium-ion batteries in SKC pumps are subject to the A-45 exemption to the IATA regulations and are not regulated for air shipments. This information must be written on the shipping document when shipping by air.

For ground shipments:
US DOT regulations specify a limit of 24 or fewer battery cells in one shipping box. If you exceed 3 QuickTake 30 pumps with battery packs in one shipping box, specific shipping requirements must be followed. Contact SKC for more information or refer to the regulatory authority in your area.

See Technical Note: SKC Sample Pump Battery Pack Characteristics and Maintenance

QuickTake 30 Operating Instructions (PDF)

QuickTake 30 Spec. Sheet (PDF)

Technical Note: SKC Sample Pump Battery Pack Characteristics and Maintenance (PDF)