SKC: Shaping the Future of Air Sampling  
 
 Español
 Deutsch
 Français
 Portuguese
 

 

 

VAC-U-GO Sampler

For Flows up to 30 L/min and Sonic Flow Applications


SKC VAC-U-GO Sampler
The Vac-U-Go non-compensating vacuum pump is a rotary vane-type pump that provides flows up to 30 L/min and maintains sonic flow when used with the SKC BioSampler ®. The AC-powered Vac-U-Go features a protective housing with handle and can be used for a variety of ambient and indoor air sampling applications including asbestos, bioaerosol, stack, VOC, and fenceline monitoring. It is ideal for sampling bioaerosols using the BioSampler ® or the AGI-30 Impinger. The SKC BioSampler does not require a critical orifice accessory for sonic flow.

The line operated Vac-U-Go Sampling pump has a permanently lubricated motor that is mounted in a housing with a vacuum pressure gauge to indicate sonic flow. and a throttling valve to adjust flow. Both models are supplied with an extension rod for mounting media and a power cord.

The Vac-U-Go Sampler can be operated in two different modes: non-compensating and BioSampler. Vac-U-Go is a non-compensating, therefore, it is not suitable for compliance sampling in non-sonic mode. For indoor air studies, compliance, or other applications that require a quiet, compensating vacuum pump, use the QuickTake 30 or Quicktake 15 Sample Pump.

Note: Sonic flow is attained when a vacuum pump pulls air through an orifice create a condition where the downstream pressure is 52% of the upstream pressure. At this point, air velocity through the orifice has reached the speed of sound and cannot increase, therefore the orifice becomes "critical." The fixed air velocity at the critical orifice also limits the airflow so that the condition remains constant regardless of pressure variations in the system. Airflow will remain constant at vacuum levels >15 inches Hg.

VAC-U-GO Operating Instructions (PDF)

Performance Profile

Flow range: Non compensated airflow up to 30 L/min.
Sonic flow with BioSampler (acts as a critical orifice) is approximately 12.5 L/min
Motor: Oil-less rotary vane
Vane life: Approximately 600 to 27,000 hrs depending on application
Power: Line power only
115V, AC, 60 Hz (Cat. No. 228-9605)
230V, AC, 50 Hz (Cat. No. 228-9605B)
Operating temperature range: 32 to 104 F (0 to 40 C)
Size: 5.5 x 10.5 x 6.5 ins. ( 13.9 x 26.7 x 16.5 cm)
Weight: 12 lbs (5.4 kg)
Standards: None. Pump is not intrinsically safe

 

Ordering Information      

Description
Cat. No.
VAC-U-GO Sampling Pump
Mounted in protective housing with vacuum gauge and valve to assure sonic flow performance. Supplied without orifices or rotameter. AC operation only.
115 V
Recommended for use with the SKC BioSampler. No critical orifice is needed if used with a BioSampler.
228-9605
VAC-U-GO Sampling Pump
Mounted in protective housing with vacuum gauge and valve to assure sonic flow performance. Supplied without orifices or rotameter. AC operation only; weighs 12 lbs (5.4 kg)
230 V
Recommended for use with the SKC BioSampler. No critical orifice is needed if used with a BioSampler.
228-9605B
Certificate of Compliance, optional, must be ordered when ordering pump P1501
Accessories L/min Cat. No.
Field Rotameter 4 to 50 320-440

VAC-U-GO Operating Instructions (PDF)



SKC Air Sampling Pumps

Pump Accessories (Chargers, Calibrators, Flowmeters, Parts)

This page was last modified April 17, 2008
  About SKC Terms of Use Privacy Policy Contact Us Ordering Information

This site was last modified May 15, 2008

SKC Inc.'s Quality Policy is in Compliance with IS0 9001-2000
© Copyright 1995-2008 SKC Inc