Grab Air Bag Sample Pump


Orderable Items The Grab Air fixed flow (1 L/min) economical area sample pump uses a 9-volt alkaline battery for sampling gases/vapors. The pump is ideal for use with sample bags.

Not UL Listed for intrinsic safety



  • Uses 9-volt alkaline battery — no down time for battery charging

  • Low-battery indicator

  • Fixed 1 L/min flow rate

  • Approximately 1000 liters volume on one battery

  • Minimum setup time — simply attach a bag and sample

     

 


 


 


Performance Profile

Flow Rate Approx. 1 L/min
Operation Time Approx. 1000 liters volume per battery dependent on battery quality
Battery Disposable 9-V alkaline
Pumping Action Diaphragm
Case Sturdy, lightweight plastic
Intrinsic Safety Not intended for use in explosive or hazardous locations
Dimensions 4.25 x 2.5 x 1.1 inches (10.8 x 6.3 x 2.7 cm)
Weight 6.6 ounces (186 grams)

Protect sample pump from weather when in use outdoors.


 


 


 


 
Grab Air Sample Pump
Economical Pump for Filling Bags


Bag Sampling Train With Grab Air Sample PumpThe SKC Grab Air Sample Pump is an economical choice for grab-and-go bag sampling. Grab Air operates at a fixed flow rate of 1 L/min for up to 1000 liters volume on one 9-volt battery. Simply attach a sample bag to the outlet port and turn on the pump. Simple, quick, economical — Grab Air.


Principle of Operation

The Grab Air Sampler pulls air containing the sample into the pump at approximately 1 L/min and immediately exhausts it into a sample bag connected to the pump with PTFE tubing. Sampling is complete
when the bag is properly filled. See Bag Sampling.

Bag Sampling

Sampling Tips

  • Use only PTFE tubing and a PTFE-lined septum for bag sampling to prevent sample loss.

  • Flush bags thoroughly with purified air or nitrogen before use.

  • Do not use bags to collect unstable or highly reactive compounds.

  • Do not ship sample bags by air unless the cargo cabin is pressurized.

  • When sampling for CO or CO2, use SKC Standard FlexFoil (263 and 263 Series) or FlexFoil PLUS (252 and 253 Series) Grab Bags.

  • Avoid filling any bag more than 80% of its maximum volume. Visually inspect the bag as it is filling to determine when sampling is complete.