  Home » Catalog » Pumps/Calibrators/Accessories » Air Sampling Pumps to 5 L/min » AirChek 52 (5 to 3000 ml/min)

AirChek 52
5 to 3000 ml/min


Orderable Items The AirChek 52 personal air sampling pump offers a 5 to 3000 ml/min flow range and 12+ hours run times for sampling particulates and gases/vapors. The pump is ideal for use with sorbent tubes, filter cassettes, size-selective samplers, and impingers.
The AirChek 52 air sampling pump is UL Listed for intrinsic safety.



 
Features

Flow range of 1000 to 3000 ml/min
Ideal for most personal sampling applications including sorbent tubes, filter cassettes, impingers, and size-selective samplers, and long-duration detector tubes; high flow range of 1000 to 3000 ml/min; low flow range of 5 to 500 ml/min requires low flow accessories

Constant flow compensation
Provides constant flow within ± 5% for accurate air volumes

For both high and low flow applications
Eliminates the need for separate sampling pumps for high and low ranges when using optional low flow accessories

Minimum controls for easy operation
Two pumps in one easy-to-use sampler

RFI/EMI shielded and UL Listed for intrinsic safety
Case material and unique circuitry are designed to provide shielding from radio frequencies in the range of 27 to 1000 MHz, UL Listed for intrinsic safety for Class I, Division 1 and 2, Groups A, B, C, D; Class II, Division 1 and 2, Groups E, F, G; and Class III, Temperature Code T3C; CE marked; ATEX-approved model available

Highly accurate display
Digital time display tracks sample run time in minutes up to 99,999
Indicates low battery condition or flow fault without losing sample run time
Attempts auto-restart up to 5 times from flow fault


Low flow sampling capabilities

  • All-in-One adjustable tube holder
  • Multiple Adjustable Tube Holder with Constant Pressure Controller (CPC)


Multiple-tube sampling capabilities
Optional multiple adjustable tube holders used with a CPC allows up to four tube samples to be taken simultaneously, each at different flow rates. Saves time and uses fewer pumps

Full shift run times
Provides 12+ hours at 2000 ml/min, 25 inches water back pressure

Small, compact and lightweight
Weighs only 20 ounces (567 grams); measures 5 x 3 x 1.75 inches (12.7 x 7.6 x 4.4 cm)

Simple design, smooth surface
The impact-resistant case is easy to clean and decontaminate. Includes sturdy attached belt clip, detachable quick change battery pack, and built-in accessory mount

Built-in pulsation dampener

Built-in replaceable particulate trap

Quiet operation

Replacement parts in preassembled modules



 


 


 


 
Performance Profile
Flow Range

1000 to 3000 ml/min
(5 to 500 ml/min requires All-in-One adjustable tube holder or Constant Pressure Controller [CPC] and Multiple Adjustable Tube Holder accessories)

Flow Control
Holds constant flow to ± 5% of set-point after calibration
Compensation Range 1000 ml/min up to 25 inches water back pressure
2000 ml/min up to 25 inches water back pressure
2500 ml/min up to 20 inches water back pressure
3000 ml/min up to 15 inches water back pressure

  Typical Back Pressure of Sampling Media (inches water)

Filter/Pore Size (µm)
Flow Rate (L/min)
1.0 1.5 2.0 2.5
25-mm MCE/0.8 6 9 12 15
25-mm MCE/0.45 14 22 28 35
37-mm MCE/0.8 2 3 4 5
37-mm PVC/5.0 1 1 2 2

Compare the information in this table to pump compensation range to determine appropriate applications.

Run Time

NiCad Battery: 8 hrs minimum at above compensation ranges; dependent on media used. See Table 1.
NiMH Battery: 12+ hrs minimum at 2000 ml/min and up to 25 inches water back pressure; dependent on media used. See Table 2.

Power Supply Rechargeable 4.8-V nickel-metal hydride (NiMH) battery pack, 3.3-Ah capacity or
Rechargeable 4.8-V nickel cadmium (NiCad) battery pack, 1.8-Ah capacity
A battery eliminator is available (see Ordering Information); use voids the UL Listing for intrinsic safety.
Battery Charging Time
(varies with battery capacity
and level of discharge)

6 to 8.5 hrs using PowerFlex Charger
Temperature Operating: 32 to 104 F (0 to 40 C)
Storage: -4 to 113 F (-20 to 45 C)
Charging: 50 to 113 F (10 to 45 C)
Humidity 0 to 95% non-condensing
Flow Fault

If pump is unable to compensate for longer than 15 seconds due to excessive back pressure, the pump enters flow fault. During flow fault, the pump stops, the fault icon appears on the display, and elapsed time remains on the display. Auto-restart is attempted up to 5 times. Elapsed time information will be lost on restart.

Low Battery Fault LCD displays low battery icon and pump shuts down. LCD remains on.
LCD Time Display LCD display shows elapsed pump run time in minutes up to 99,999.
LCD Indicator LCD always remains on. Displayed elapsed run time information resets when pump is started.
Noise Level

62.5 dBA* - pump without case
55 dBA* - pump housed in noise-reducing case (optional accessory Cat. No. 224-96C)

* Pump operating at 2 L/min with a 37-mm, 0.8-µm MCE filter cassette measured at 3 ft (1 m) from pump
RFI/EMI Shielding Complies with requirements of EN 55022 1987 Class
B, and EN 50082-1:1992
Frequency range of radiated susceptibility test was
27 to 1000 MHz.
CE marked
Intrinsic Safety UL Listed for Class I, Division 1 and 2, Groups A, B, C, D; Class II, Division 1 and 2, Groups E, F, G; and, Class III, Temperature Code T3C. ATEX-approved models available; contact SKC
Dimensions 5 x 3 x 1.75 in (12.7 x 7.6 x 4.4 cm)
Weight 20 oz (567 gm)
Multiple-tube Sampling Adjustable tube holder accessory allows user to take up to four simultaneous samples at flows from 5 to 500 ml/min. CPC required
Tubing Requires 1/4-in ID tubing

Protect sample pump from weather when in use outdoors.
Weather enclosures for outdoor pump use are available. Contact SKC
.


 


 


 


 
AirChek 52 Low Flow Capabilities
The full range of flow rates is available in
one lightweight, low-cost, heavy-duty sampler.

The AirChek 52 Sample Pump is a fully versatile sampler that can be used for low flow (5 to 500 ml/min) sampling. Use the pump with the All-in-One Adjustable Tube Holder with built-in Constant Pressure Controller (CPC) for single-tube sampling or with a multiple Adjustable Tube Holder and the CPC accessory (Cat. No. 226-26-CPC) for multiple tube sampling.

All-in-One Low Flow Holder connected to Airchek 52

For multiple tubes: Simple to use, the CPC fits into the sampling train tubing, allowing the pump to operate in a constant pressure mode. In the low flow mode, a multiple adjustable tube holder can be used to collect up to four tube samples simultaneously and at different flow rates. The flow rate at each intake of the adjustable tube holders can be set from 5 to 500 ml/min (dependent on back pressure). Multiple-tube sampling reduces time in the field and uses fewer pumps.


Lightweight, Durable


Measuring approximately 5 x 3 x 1.75 inches (12.7 x 7.6 x 4.4 cm) and weighing only 20 ounces (567 grams), the small size, light weight, and durable construction allows the AirChek 52 to withstand the rigors of on-worker use time and time again. Rechargeable NiMH or NiCad batteries provide a reliable power source which can be restored overnight. The AirChek 52 is the lightweight, inexpensive workhorse pump designed for rugged and repeated use. Now the full range of flow rates is available in one lightweight, low-cost, heavy-duty sampler to use with:
  • Sorbent sample tubes and Long-duration detector tubes for collecting most hazardous chemical gases and vapors from the air
  • Cyclones for collecting respirable dust
  • IOM samplers for collecting inhalable dust
  • Filters for collecting total dust, asbestos, or chemical hazards
  • Impingers for collecting reactive chemicals into liquids

 

 

Table 1. AirChek 52 Run Time in Hours with NiCad Battery

Following are typical run times achieved when using a fully charged nickel cadmium (NiCad) battery pack. Data is sorted by type of sample media. All run times are listed in hours.

Mixed Cellulose Ester (MCE) filter, 0.8-µm pore size

 
Filter Diameter
Flow Rate (L/min)
37 mm
25 mm
2.0
24.5
16.5
2.5
22.9
12.7
3.0
18.9
**

Polyvinyl Chloride (PVC) filter, 5.0-µm pore size

 
Filter Diameter
Flow Rate (L/min)
37 mm
25 mm
2.0
25.8
24.9
2.5
24.3
21.1
3.0
25.1
18.6
 ** Filter back pressure exceeded pump capability during testing.
Note: Increases in back pressure during sampling due to buildup of sample on the filter can decrease battery life.

 

Table 2. AirChek 52 Run Time in Hours with NiMH Battery

Following are typical run times achieved when using a fully charged nickel-metal hydride (NiMH) battery pack. Data is sorted by type of sample media. All run times are listed in hours. Results are obtained using a new pump and new fully charged battery. Pump performance may vary.

Mixed Cellulose Ester (MCE) filter, 0.8-µm pore size

 
Filter Diameter
Flow Rate (L/min)
37 mm
25 mm
2.0
40
37
2.5
38
29
3.0
34
25

Polyvinyl Chloride (PVC) filter, 5.0-µm pore size

 
Filter Diameter
Flow Rate (L/min)
37 mm
25 mm
2.0
56
47
2.5
48
41
3.0
42
38

Note: Increases in back pressure during sampling due to buildup of sample on the filter can decrease battery life.