1. Calibration: How do you calibrate samplers like cyclones and similar size-selective samplers?

Calibration Option 1- Preferred OptionThe most accurate calibration procedure is to use the calibration adapter designed specifically for the sampler. The pump should be connected to the suction port of a calibrator with the sampler/media/cal adapter assembly…